The LNI Schmidlin PermaCal
Calibration Gas Generator allows the preparation of Calibration Gas
Standards at a wide user-definable concentration range.
PermaCal uses an external source of
Ultra Zero Air and gas permeation tubes, combined with mass flow
controllers to produce user-definable gas calibration standards.
A gas permeation tube contains a specific substance in liquid and
vapour phases. The diffusion rate of the vapour phase through
the tube is a function of temperature. By maintaining the tube
at a constant controlled temperature, the gas permeation rate
becomes constant. Using mass flow controllers to generate a
range of dilution flows of Ultra Zero Air, PermaCal can generate
specific concentrations of calibration gas.
Applications:
Features:
Permeation oven: accepts up to 6 permeation tubes (up to 12 cm long)
Oven Temperature: 40 - 80 oC
Permeation oven flow: 100 mL/min
Total flow: 500 to 5,000 mL/min (depending on concentration setting)
Requires external source of Ultra Zero Air
Inlet pressure: 45 psig
Outlet pressure: atmospheric pressure
Small size (6" W x 14" D x 15" H) conserves lab bench space
LCD touch-screen display provides simplified user interface
Weight: 31 lbs
1 year parts & labour warranty
CE & CSA certified